9 CHECK AND ADJUSTMENT
9.7 How to Set the Instrument Constant Value
To set the Instrument constant which is obtained in section 9.1 “Checking and Adjusting of Instrument
Constant”, follow as below.
Operating procedure Operation Display
1
Press the [F5] key from the main menu.
[F5]
2
Press [F4] (Instrument constant) key.
[F4]
3
Press [F5] (YES) key .
[F5]
4
Input value and press [ENT] key.
Input
value
[ENT]
5
Press [F5](OK) key.
The display returns to main menu.
[F5]
Adjustment
F1 V0/Axis (Measurement)
F2 V0/Axis (Constant list)
F3 Date Time
F4 Instrument constant
Instrument Constant
EDM OFSET (mm)
0.0
Modify
YES NO
Instrument Constant
EDM OFFSET (mm)
0
.0
EXIT BS
Instrument Constant
EDM OFFSET (mm)
1.2
OK CANCEL
Complete